Laureate™ True RMS Voltage & Current
Meter | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| AC Voltage FS Range | AC Voltage to Overrange Flash | Resolution | Input Resistance | Error at 25°C |
| 200.00 mV 2.0000 V 20.000 V 200.00 V |
264.00 mV 2.640 V 26.400 V 264.00 V |
10 µV 100 µV 1 mV 10 mV |
1 MΩ 1 MΩ 1 MΩ 1 MΩ |
± 0.1% of FS, 0-100% of FS, 10 Hz to 10 kHz (AC coupling) or DC to 10 kHz (DC coupling) |
| 600.0 V * | 650 V | 100 mV | 1 MΩ | ± 2V |
| AC Current FS Range | AC Current to OverrangeFlash | Resolution | Input Resistance | Error at 25°C ± 2 cts |
| 2.0000 mA 20.000 mA 200.00 mA |
2.6400 mA 26.400 mA 264.00 mA |
0.1 µA 1.0 µA 10 µA |
100 Ω 10 Ω 1 Ω |
± 0.1% of FS, 0-100% of FS, 10 Hz to 10 kHz (AC coupling) or DC to 10 kHz (DC coupling) |
| 5.000 A | 5.4 A | 1 mA | 0.01 Ω | ± 20 mA |
| Display | |
| Readout Range Display Update Rate Indicators Crest factor Vp / Vrms |
5 LED digits, 7-segment, 14.2 mm (.56"), red or green -99999 to 99999 or -99990 to 99990 (count by 10) 3.75/s at 60 Hz power, 3.1/s at 50 Hz power 1 LED lamp per relay setpoint 3.00 at full scale range, 2.27 at 132% of FS range (display flashes), 1.414 at 212% of FS range (maximum sinusoidal signal) |
| Output Update Rate | |
| A-to-D rate Signals > 50/60 Hz Signals 3 to 50/60 Hz Signals DC to 3 Hz |
60/s at 60 Hz power, 50/s at 50 Hz power 60/s at 60 Hz power, 50/s at 50 Hz power Signal frequency 3 per second |
| Maximum Signal | |
| Max applied voltage Current protection |
600 Vac for 2, 20, 200, 600V ranges, 50 Vac for 0.2V range 25x for 2 mA, 8x for 20 mA, 2.5x for 200 mA, 1x for 5A |
| Power | |
| Voltage, standard Voltage, optional Frequency Power Isolation |
85-264 Vac or 90-300 Vdc (DC operation not UL approved) 12-32 Vac or 10-48 Vdc DC or 47-63 Hz 250V rms working, 2.3 kV rms per 1 min test |
| Analog Output (optional) | |
| Output Levels Current compliance Voltage compliance Scaling Resolution Isolation |
4-20 mA, 0-20 mA, 0-10V, -10 to +10V (jumper selectable) 2 mA at 10V ( > 5 kΩ load) 12V at 20 mA ( < 600Ω load) Zero and full scale adjustable from -99999 to +99999 16 bits (0.0015% of full scale) 250V rms working, 2.3 kV rms per 1 min test |
| Relay Outputs (optional) | |
| Relay Types Current Ratings Output common Isolation |
2 Form C contact relays or 4 Form A contact relays (NO) 2 or 4 Form A, AC/DC solid state relays (NO) 8A at 250 Vac or 24 Vdc for contact relays 130 mA at 140 Vac or 180 Vdc for solid state relays Isolated commons for dual relays or each pair of quad relays 250V rms working, 2.3 kV rms per 1 min test |
| Serial Data I/O (optional) | |
| Board Selections Protocols Data Rates Digital Addresses Isolation |
Ethernet, Ethernet-to-RS485 server, USB, USB-to-RS485 server,
RS485 (dual RJ11), RS485 Modbus (dual RJ45), RS232. Modbus RTU, Modbus ASCII, Laurel ASCII protocol 300 to 19200 baud 247 (Modbus), 31 (Laurel ASCII), 250V rms working, 2.3 kV rms per 1 min test |
| Signal Connections | |
![]() |
|
| Environmental | |
| Operating Temperature Storage Temperature Relative Humidity Protection |
0°C to 60°C -40°C to 85°C 95% at 40°C, non-condensing NEMA-4X (IP-65) when panel mounted |