MAGNA Series, Strain Gauge / Load Cell Input Large Digit Displays

For Use with Strain Gauges & Load Cells. Includes Excitation for Four 350-Ohm Load Cells in Parallel.
Approvals
Types Produced to IPC 610 Class 2 Standard
UL ISO9001:2008 QA
Strain Gauge / Load Cell Input
A/D Converter
Signal Gain
Input Resistance
Accuracy
Tempco, Scale
Tempco, Offset
Excitation Output
20-bit resolution
Suitable for 0.5 mV / V to 4 mV / V load cells
> 10 MΩ
± 0.05% of range.
± 25 ppm/°C
± 30 ppm/°C
Isolated 10V up to 120 mA to power up to four 350Ω load cells in parallel
Display
Number of Digits
Digit Height
Digit Color
Brightness
Numeric Format
Display Update Rate
4 or 6
57 mm (2-1/4"), 102 mm (4.0"), 150 mm (5.9"), or 200 mm (7.9")
Red
Variable
8.8.8.8. or 8.8.8.8.8.8.
10 readings/second
User Controls
Scaling
Filtering
Rounding
Linearization
Digital direct or theoretical scaling with scale & offset. 
Digital filtering, time constants adjustable from 0 to 25 sec. 
Selectable last-digit rounding of 1, 2, 5, 10, 20, 50 units.
Up to 10 points, direct or theoretical calibration.
Logic Inputs
Front Panel Controls
Logic Functions
  
Logic Levels
Four buttons to set time, date, alarms, etc.
Three logic inputs, individually configurable for Hold, Reset, Tare, Peak / Valley, Net/Gross, Memory Address.
Normally open, NPN or contact closure input to activate.
5V open circuit, 1 mA switched current.
Power
AC Power (standard)
DC Power (option)
Max Consumption
100-240 Vac
11-30 Vdc
30 VA
Analog Output Option
Output Levels
  
  
Resolution
Accuracy
Stability
Output Isolation
Scaling
Special Features
0-20 / 4-20 mA into 0 to 500Ω, 0.4 µA resolution.
0-10V into loads > 600Ω, 0.2 mV resolution.
-10 to+10V into loads > 600Ω, 0.4 mV resolution.
16 bits
0.1% of range
50 ppm/°C
Isolated from input and power
Digitally scalable
Forward or reverse action possible.
Can be set to operate from net or gross value.
Alarm Output Option
Relay Type
Special Features
2 Alarms rated 5A, 250VAC, SPDT
Adjustable hysteresis
In-flight compensation facility.
Energize or de-energize on trip.
Adjustable timers to set energize and de-energize delays. Can also be set to pulse for applications needing less than 100% duty cycle.
In-band and out-of-band alarm function.
Manual and automatic in-flight compensation.
Can be set to operate from net or gross value.
Serial Data Output Option
Output Types
Special Features
RS232 or RS485, addressable, on demand or continuous output.
Can include time and date with RTC option fitted.
Can be set to operate from net or gross value.
Environmental
Storage Temp.
Operating Temp.
Sealing
Mounting
-20°C to +70°C
0C to 50°C, non-condensing
IP65 standard, all-round
Wall mount, suspension mount, panel mount