Laureate™ AC Power Factor Meter & AC Phase Angle Meter

Navigation:  Home > Electronic Counters
AC phase angle & power factor meter
Key Features
  • Minus sign for power factor with negative phase angle
  • Phase angle resolution of 1÷, 0.1÷ or 0.01÷
  • Phase angle accuracy of 0.05÷ at AC line frequency
  • AC line voltages up to 250 Vrms
  • AC frequency of 1 Hz to 10 kHz
  • Universal AC input power, 85-264 Vac
AC Power Factor Meter & AC Phase Angle Meter
Phase Angle Mode
Item Displayed
Display Units
Frequency Range
Accuracy
Phase angle difference between two waves of same period
1°, 0.1°, 0.01°
0.005 Hz to 10 kHz
0.01°, 1 Hz to 100 Hz; 0.1° at 1 kHz; 1° at 10 kHz
Power Factor Mode
Item Displayed
Display Units
Polarity
Frequency Range
Accuracy
Power factor between two sine waves of same period
1.000 to 0.000, 1.00 to 0.00, or 1.0 to 1.0
Negative sign indicates negative phase angle
0.005 Hz to 10 kHz
0.1% at power line frequencies
Update Rate
Timing interval
Gate Time
Time Before Zero Out
Gate time + 30 ms+ 0-2 signal periods
Selectable 10 ms to 199.99 s
Selectable 10 ms to 199.99 s (to indicate loss of signal)
Display
Readout
Range
Indicators
6 LED digits, 7-segment, 14.2 mm (.56"), red or green
-999999 to +999999
Four LED lamps
Inputs
Signal ranges
Signal ground
Noise filter
Nine AC signal ranges from 12 mVp-p to 250 Vac
Common ground for channels A & B
1 MHz, 30 kHz, 250 Hz (jumper selectable) + digital filter
Power
Voltage, standard
Voltage, optional
Frequency
Power Isolation
85-264 Vac or 90-300 Vdc (DC operation not UL approved)
12-32 Vac or 10-48 Vdc
DC or 47-63 Hz
250V rms working, 2.3 kV rms per 1 min test
Excitation Output (standard)
5 Vdc
10 Vdc
24 Vdc
Output Isolation
5 Vdc ± 5%, 100 mA
10 Vdc ± 5%, 120 mA
24 Vdc ±5 %, 50 mA
50 Vdc to meter ground
Analog Output (optional)
Output Levels
Current compliance
Voltage compliance
Scaling
Resolution
Isolation
4-20 mA, 0-20 mA, 0-10V, -10 to +10V (jumper selectable)
2 mA at 10V ( > 5 kΩ load)
12V at 20 mA ( < 600Ω load)
Zero and full scale adjustable from -99999 to +99999
16 bits (0.0015% of full scale)
250V rms working, 2.3 kV rms per 1 min test
Relay Outputs (optional)
Relay Types

Current Ratings

Output common
Isolation
2 Form C contact relays or 4 Form A contact relays (NO)
2 or 4 Form A, AC/DC solid state relays (NO)
8A at 250 Vac or 24 Vdc for contact relays
130 mA at 140 Vac or 180 Vdc for solid state relays
Isolated commons for dual relays or each pair of quad relays
250V rms working, 2.3 kV rms per 1 min test
Serial Data I/O (optional)
Board Selections

Protocols
Data Rates
Digital Addresses
Isolation
EthernetEthernet-to-RS485 serverUSBUSB-to-RS485 server,
RS485 (dual RJ11), RS485 Modbus (dual RJ45), RS232.
Modbus RTU, Modbus ASCII, Laurel ASCII protocol
300 to 19200 baud
247 (Modbus), 31 (Laurel ASCII), 
250V rms working, 2.3 kV rms per 1 min test
Environmental
Operating Temperature
Storage Temperature
Relative Humidity
Protection
0°C to 60°C
-40°C to 85°C
95% at 40°C, non-condensing
NEMA-4X (IP-65) when panel mounted
Signal Connections
Signal connections to Laurel pulse input batch controller
feedback